To carry out operations for the installation of dental implants, a set from Saardent`s Implants is required. The accessories of the surgical kit are placed on the box panel to facilitate the work of the implantologist as much as possible. They will help to reduce the time of the operation, make it less traumatic, and prevent possible complications. The special design of the implant driver allows maintaining the sterility of the implant and facilitates its manipulation. In the set:
- Long and short implant drivers for the surgical handpiece. Provide easy extraction and fixation of the implant, and then its placement in the bed;
- Screwdriver wrench for screws – long and short;
- Parallel pins – visualize the insertion path and parallelism during drilling;
- Adapter for implant driver – ratchet adapter;
- Ratchet wrench – used to tighten the implant and screws;
- Pilot drill – used for primary drilling;
- Conical and expansion drills of different diameters allow accurate and safe formation of the stock, thanks to the special markings on the drills, constant monitoring of the operation is carried out.
The surgical kit is packaged in an ergonomic box. As a “tip” to the surgeon, the instruments are arranged in multi-colored cells and are marked.
- Wrench Ratchet – ratchet wrench
- Pilot Drill Ø 1.5 – pilot drill 1.5
- Start Drill Ø 2.0 – Starter drill 2.0
- Start Drill Ø 2.8 – starter drill 2.8
- Conical drill Ø 3.45 – conical drill 3.45
- Conical drill Ø 3.9 – conical drill 3.9
- Conical drill Ø 4.3 – conical drill 4.3
- Parallel pin – parallel pin
- Parallel pin – parallel pin
- Hand Driver Hex. 1.20 9mm
Screwdriver key for screws 1.20 9mm - Hand Driver Hex. 1.20 18 mm
Screwdriver key for screws 1.20 18 mm - Implantdriver for Hand Piece 9 mm
Implant driver for surgical handpiece 9 mm - Implantdriver for Hand Piece 18 mm
Implant driver for surgical handpiece 18 mm - Driver Adapter – adapter for implant driver
Pilot drill 1.5 | Starter drill 2.0 | Starter drill 2.8 |
PD15 | SD20 | SD28 |
Conical drill 3.45 | Conical drill 3.9 | Conical drill 4.3 |
CD345 | CD39 | CD43 |
Screwdriver key for screws 1.20 9 mm | Screwdriver key for screws 1.20 18 mm | Implant driver adapter |
HD Hex. 1.20 – 9 | HD Hex. 1.20 – 18 | DA |
Implant driver for surgical handpiece 18 mm | Implant driver for surgical handpiece 9 mm | Parallel pin |
IHP18 | IHP9 | PP |
Ratchet wrench |
WR |
Torque adapter 10-45 Ncm |
RT |
* Torque adapter included in advanced surgical kit
Surgical protocol
Sequence of use of cutters
- By adhering to the drilling protocol, the Dentinium implants can achieve optimum primary stability.
- The Dentinium implant concept implies subcrestal immersion of the implant 0.5 mm.
Implant placement protocol Ø 3.5 *
* When installing an implant with a length of 10 mm.
Implant placement protocol Ø 4.0 *
* When installing an implant with a length of 10 mm.
Implant placement protocol Ø 4.5 *
* When installing an implant with a length of 10 mm.
Surgical drill protocol
Drill diameter, mm | 1.5 | 2.0 | 2.8 | 3.45 | 3.9 | 4.3 |
Speed drilling,rpm | D1 800-1200D2 800-1200D3 800-1000D4 800-1000 | D1 800-1000D2 800-1000D3 700-900D4 600-800 | D1 600-800D2 600-800D3 500-600D4 400-500 | D1 500-700D2 600-800D3 500-600D4 400-500 | D1 400-600D2 500-700D3 500-600D4 400-500 | D1 400-600D2 500-700D3 500-600D4 400-500 |
Subcrestal implant placement
The implant immersion depth is easily controlled due to the straight shape of the implant neck, which minimizes bone tissue atrophy.
Attention! The actual length of the Dentinium implants is 0.5 mm less than the stated length, which ensures the immersion of the implants below the level of the cortical layer.