Surgical Kit

To carry out operations for the installation of dental implants, a set from Saardent`s Implants is required. The accessories of the surgical kit are placed on the box panel to facilitate the work of the implantologist as much as possible. They will help to reduce the time of the operation, make it less traumatic, and prevent possible complications. The special design of the implant driver allows maintaining the sterility of the implant and facilitates its manipulation. In the set:

  • Long and short implant drivers for the surgical handpiece. Provide easy extraction and fixation of the implant, and then its placement in the bed;
  • Screwdriver wrench for screws – long and short;
  • Parallel pins – visualize the insertion path and parallelism during drilling;
  • Adapter for implant driver – ratchet adapter;
  • Ratchet wrench – used to tighten the implant and screws;
  • Pilot drill – used for primary drilling;
  • Conical and expansion drills of different diameters allow accurate and safe formation of the stock, thanks to the special markings on the drills, constant monitoring of the operation is carried out.

The surgical kit is packaged in an ergonomic box. As a “tip” to the surgeon, the instruments are arranged in multi-colored cells and are marked.

  1. Wrench Ratchet – ratchet wrench
  2. Pilot Drill Ø 1.5 – pilot drill 1.5
  3. Start Drill Ø 2.0 – Starter drill 2.0
  4. Start Drill Ø 2.8 – starter drill 2.8
  5. Conical drill Ø 3.45 – conical drill 3.45
  6. Conical drill Ø 3.9 – conical drill 3.9
  7. Conical drill Ø 4.3 – conical drill 4.3
  8. Parallel pin – parallel pin
  9. Parallel pin – parallel pin
  10. Hand Driver Hex. 1.20 9mm
    Screwdriver key for screws 1.20 9mm
  11. Hand Driver Hex. 1.20 18 mm
    Screwdriver key for screws 1.20 18 mm
  12. Implantdriver for Hand Piece 9 mm
    Implant driver for surgical handpiece 9 mm
  13. Implantdriver for Hand Piece 18 mm
    Implant driver for surgical handpiece 18 mm
  14. Driver Adapter – adapter for implant driver
Pilot drill 1.5Starter drill 2.0Starter drill 2.8
PD15SD20SD28
Conical drill 3.45Conical drill 3.9Conical drill 4.3
CD345CD39CD43
Screwdriver key for screws 1.20 9 mmScrewdriver key for screws 1.20 18 mmImplant driver adapter
HD Hex. 1.20 – 9HD Hex. 1.20 – 18DA
Implant driver for surgical handpiece 18 mmImplant driver for surgical handpiece 9 mmParallel pin
IHP18IHP9PP
Ratchet wrench
WR
Torque adapter 10-45 Ncm
RT

* Torque adapter included in advanced surgical kit

Surgical protocol

Sequence of use of cutters

  • By adhering to the drilling protocol, the Dentinium implants can achieve optimum primary stability.
  • The Dentinium implant concept implies subcrestal immersion of the implant 0.5 mm.

Implant placement protocol Ø 3.5 *

* When installing an implant with a length of 10 mm.

Implant placement protocol Ø 4.0 *

* When installing an implant with a length of 10 mm.

Implant placement protocol Ø 4.5 *

* When installing an implant with a length of 10 mm.

Surgical drill protocol

Drill diameter, mm1.52.02.83.453.94.3

Speed drilling,rpm

D1 800-1200D2 800-1200D3 800-1000D4 800-1000

D1 800-1000D2 800-1000D3 700-900D4 600-800

D1 600-800D2 600-800D3 500-600D4 400-500

D1 500-700D2 600-800D3 500-600D4 400-500

D1 400-600D2 500-700D3 500-600D4 400-500

D1 400-600D2 500-700D3 500-600D4 400-500

Subcrestal implant placement

The implant immersion depth is easily controlled due to the straight shape of the implant neck, which minimizes bone tissue atrophy.

Attention! The actual length of the Dentinium implants is 0.5 mm less than the stated length, which ensures the immersion of the implants below the level of the cortical layer.